For each citation that was shared on social media (LinkedIn, Facebook, or Twitter) with the “@GenScript” tag, the author will be rewarded with a $10 Amazon gift card or 2,000 GS points.

Performance characteristics of the VIDAS® 25-OH Vitamin D Total assay - comparison with four immunoassays and two liquid chromatography-tandem mass spectrometry methods in a multicentric study

Clin Chem Lab Med. 2015; 
Moreau E, Bächer S, Mery S, Le Goff C, Piga N, Vogeser M, Hausmann M, Cavalier E.
Products/Services Used Details Operation
Peptide Synthesis … Unlabeled and labeled peptides were purchased from GenScript (Piscataway, NJ), EZBiolab (Carmel, IN) or Biomatik (Wilmington, DE) SRM 1949 Frozen Human Prenatal Serum was from NIST (Gaithersburg, MD) 22 Peptide calibrant preparation … Get A Quote

Abstract

The study was conducted to evaluate the analytical and clinical performance of the VIDAS® 25-OH Vitamin D Total assay. The clinical performance of the assay was compared with four other immunoassays against the results of two different liquid chromatography/mass spectrometry methods (LC-MS/MS) standardized to NIST reference materials.,VIDAS® 25-OH Vitamin D Total assay precision, linearity, detection limits and sample matrix comparison were assessed following CLSI guidelines. For method comparison, a total of 150 serum samples ranging from 7 to 92 ng/mL were analyzed by all the methods. Correlation was studied using Passing-Bablok regression and Bland-Altman analysis. The concordance correlation coefficient (... More

Keywords